Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6…
High7.5EPSS 1.0%
- Published
- Updated
- Reserved
- Assigned by
- icscert
Description
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| NVDNVD · Modified | 3.0 | High7.5 | CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
- EPSS (exploitation in 30 days)
- 1.0% · top 37% of all CVEs (4 Nov 2026)
Loading the EPSS history…
The list uses the assigner's score when there is one, then CISA's, then NVD's: High 7.5.
Weakness
Affected products
Hospira · Plum A+ Infusion System
- affected <= 13.4
Hospira · Plum A+3 Infusion System
- affected <= 13.6
Hospira · Symbiq Infusion System
- affected <= 3.13
References (1)
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 (opens in a new tab)x_refsource_MISCx_transferred
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket
[](https://trackcve.app/cve/CVE-2015-3952)Record last updated .