CVE-2017-12714

Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up"…

Medium6.5EPSS 0.63%
Published
Updated
Reserved
Assigned by
icscert

Description

Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

Scores

CVSS scores of CVE-2017-12714, by who scored it
Scored byCVSSScoreVector
NVDNVD · Modified3.0Medium6.5CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
EPSS (exploitation in 30 days)
0.63% · top 49% of all CVEs (5 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 6.5.