Skip to content
CVE Tracker
  • Newest CVEsPage
  • Recently added to KEVPage
  • Rising exploitation (EPSS)Page
  • Weekly digestPage
  • StatisticsPage
  • WatchlistPage
  • API documentationPage
  • Data sourcesPage
  • ChangelogPage
  • AboutPage
  • How it's builtPage
Back

CVE-2017-12718

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.…

High8.1EPSS 12.8%
Published
Updated
Reserved
Assigned by
icscert

Description

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.

Scores

CVSS scores of CVE-2017-12718, by who scored it
Scored byCVSSScoreVector
NVDNVD · Modified3.0High8.1CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS (exploitation in 30 days)
12.8% · top 3.7% of all CVEs (4 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: High 8.1.

Weakness

Affected products

  • Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

    • affected Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

References (4)

Sources

The same CVE in the primary databases and aggregators, to check against.

Badge for a README or a ticket

CVE-2017-12718 badge
[![CVE-2017-12718](https://trackcve.app/badge/CVE-2017-12718.svg)](https://trackcve.app/cve/CVE-2017-12718)

Record last updated .