CVE-2017-12720

An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and…

High8.1EPSS 1.9%
Published
Updated
Reserved
Assigned by
icscert

Description

An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not require authentication if the pump is configured to allow FTP connections.

Scores

CVSS scores of CVE-2017-12720, by who scored it
Scored byCVSSScoreVector
NVDNVD · Modified3.0High8.1CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS (exploitation in 30 days)
1.9% · top 21% of all CVEs (5 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: High 8.1.

Weakness

Affected products

  • Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

    • affected Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump