CVE-2017-12726

A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and…

High7.3EPSS 1.1%
Published
Updated
Reserved
Assigned by
icscert

Description

A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external communications. Smiths Medical assesses that it is not possible to upload files via Telnet and the impact of this vulnerability is limited to the communications module.

Scores

CVSS scores of CVE-2017-12726, by who scored it
Scored byCVSSScoreVector
NVDNVD · Modified3.0High7.3CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L
EPSS (exploitation in 30 days)
1.1% · top 35% of all CVEs (5 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: High 7.3.

Weakness

Affected products

  • Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

    • affected Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump