CVE-2017-13993
An Uncontrolled Search Path or Element issue was discovered in i-SENS SmartLog Diabetes Management Software, Version 2.4.0 and prior…
- Published
- Updated
- Reserved
- Assigned by
- icscert
Description
An Uncontrolled Search Path or Element issue was discovered in i-SENS SmartLog Diabetes Management Software, Version 2.4.0 and prior versions. An uncontrolled search path element vulnerability has been identified which could be exploited by placing a specially crafted DLL file in the search path. If the malicious DLL is loaded prior to the valid DLL, an attacker could execute arbitrary code on the system. This vulnerability does not affect the connected blood glucose monitor and would not impact delivery of therapy to the patient.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| NVDNVD · Modified | 3.0 | High7.8 | CVSS:3.0/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 1.8% · top 22% of all CVEs (3 Nov 2026)
Loading the EPSS history…
The list uses the assigner's score when there is one, then CISA's, then NVD's: High 7.8.
Weakness
Affected products
i-SENS, Inc. SmartLog Diabetes Management Software
- affected i-SENS, Inc. SmartLog Diabetes Management Software
References (2)
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-01 (opens in a new tab)x_refsource_MISCx_transferred
- 100659 (opens in a new tab)vdb-entryx_refsource_BIDx_transferred
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Record last updated .