Skip to content
CVE Tracker
  • Newest CVEsPage
  • Recently added to KEVPage
  • Rising exploitation (EPSS)Page
  • Weekly digestPage
  • StatisticsPage
  • WatchlistPage
  • API documentationPage
  • MCP serverPage
  • Data sourcesPage
  • ChangelogPage
  • AboutPage
  • How it's builtPage
Back

CVE-2018-10634

Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Cleartext Transmission of Sensitive Information

Medium4.8EPSS 0.47%
Published
Updated
Reserved
Assigned by
icscert

Description

Communications between Medtronic MiniMed MMT pumps and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.

Timeline

  1. ID reserved

    By icscert

  2. · 3 months later

    Published

  3. · 7 years later

    Record updated

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2018-10634, by who scored it
Scored byCVSSScoreVector
icscertCNA (assigner)3.1Medium4.8CVSS:3.1/AV:A/AC:H/PR:N/UI:R/S:U/C:H/I:N/A:N
NVDNVD · Modified3.1Medium4.8CVSS:3.1/AV:A/AC:H/PR:N/UI:R/S:U/C:H/I:N/A:N
EPSS (exploitation in 30 days)
0.47% · top 59% of all CVEs (7 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 4.8.

Weakness

Affected products

What to do

Workaround: The remote option is turned off in the pump by default.   Medtronic is directing all users to stop using their remote controllers, disable the remote option on their insulin pump, and to return the remote controllers to Medtronic. Medtronic has released additional patient focused information https://www.medtronic.com/security . Additionally, Medtronic will be sending a letter to patients who may still be actively using the remotes in order to inform patients about these security risks, and request patients stop using the remote and return them to Medtronic.

References (3)

Sources

The same CVE in the primary databases and aggregators, to check against.

Badge for a README or a ticket

CVE-2018-10634 badge
[![CVE-2018-10634](https://trackcve.app/badge/CVE-2018-10634.svg)](https://trackcve.app/cve/CVE-2018-10634)