Skip to content
CVE Tracker
  • Newest CVEsPage
  • Recently added to KEVPage
  • Rising exploitation (EPSS)Page
  • Weekly digestPage
  • StatisticsPage
  • WatchlistPage
  • API documentationPage
  • MCP serverPage
  • Data sourcesPage
  • ChangelogPage
  • AboutPage
  • How it's builtPage
Back

CVE-2018-8861

Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6…

High8.7EPSS 0.40%
Published
Updated
Reserved
Assigned by
icscert

Description

Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) could enable a limited-access kiosk user or an unauthorized attacker to break-out from the containment of the kiosk environment, attain elevated privileges from the underlying Windows OS, and access unauthorized resources from the operating system.

Scores

CVSS scores of CVE-2018-8861, by who scored it
Scored byCVSSScoreVector
NVDNVD · Modified3.0High8.7CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:L
EPSS (exploitation in 30 days)
0.40% · top 66% of all CVEs (5 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: High 8.7.

Weakness

Affected products

  • Philips · Brilliance CT Scanners

    • affected Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior.

References (3)

Sources

The same CVE in the primary databases and aggregators, to check against.

Badge for a README or a ticket

CVE-2018-8861 badge
[![CVE-2018-8861](https://trackcve.app/badge/CVE-2018-8861.svg)](https://trackcve.app/cve/CVE-2018-8861)

Record last updated .