Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

Medium6.2EPSS 0.33%
Published
Updated
Reserved
Assigned by
icscert

Description

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Scores

CVSS scores of CVE-2018-8868, by who scored it
Scored byCVSSScoreVector
icscertCNA (assigner)3.1Medium6.2CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L
NVDNVD · Modified3.1Medium6.2CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L
EPSS (exploitation in 30 days)
0.33% · top 73% of all CVEs (4 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 6.2.

Weakness

Affected products

What to do

Solution: Medtronic will release several rolling over-the-air product updates that will mitigate the vulnerabilities described within this advisory. These updates will be applied to devices automatically as part of standard, reoccurring update processes. In addition, Medtronic has increased security monitoring of affected devices and related infrastructure. Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security

Workaround: Medtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: * Maintain good physical controls over the home monitor as the best mitigation to these vulnerabilities.   * Only use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system. * Report any concerning behavior regarding their home monitor to their healthcare provider or a Medtronic representative. Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security