CVE-2019-13535

Medtronic Valleylab FT10 and LS10 Protection Mechanism Failure

Medium4.6EPSS 0.37%
Published
Updated
Reserved
Assigned by
icscert

Description

In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.

Scores

CVSS scores of CVE-2019-13535, by who scored it
Scored byCVSSScoreVector
icscertCNA (assigner)3.1Medium4.6CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
NVDNVD · Modified3.1Medium4.6CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS (exploitation in 30 days)
0.37% · top 69% of all CVEs (6 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 4.6.

Weakness

Affected products

What to do

Solution: A software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact https://www.medtronic.com/security