Medtronic Valleylab FT10 and FX8 Reversible One-way Hash

High7.0EPSS 0.27%
Published
Updated
Reserved
Assigned by
icscert

Description

Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use the descrypt algorithm for OS password hashing. While interactive, network-based logons are disabled, and attackers can use the other vulnerabilities within this report to obtain local shell access and access these hashes.

Timeline

  1. ID reserved

    By icscert

  2. · 4 months later

    Published

  3. · 6 years later

    Record updated

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2019-13539, by who scored it
Scored byCVSSScoreVector
icscertCNA (assigner)3.1High7.0CVSS:3.1/AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:H
NVDNVD · Modified3.1High7.8CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
EPSS (exploitation in 30 days)
0.27% · top 81% of all CVEs (7 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: High 7.0.

Weakness

Affected products

What to do

Solution: Software patches are currently available for the FT10 platform and will be available in early 2020 for the FX8 platform. Until these updates can be applied, Medtronic recommends to either disconnect affected products from IP networks or to segregate those networks, such that the devices are not accessible from an untrusted network (e.g., Internet). Patches can be downloaded at the following location: https://www.medtronic.com/covidien/en-us/support/software.html Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security