Back
CVE-2019-5024
A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical…
High7.6EPSS 0.47%
- Published
- Updated
- Reserved
- Assigned by
- talos
Description
A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| talosCNA (assigner) | 3.0 | High7.6 | CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
| NVDNVD · Modified | 3.1 | High7.6 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 0.47% · top 59% of all CVEs (6 Nov 2026)
Loading the EPSS history…
The list uses the assigner's score when there is one, then CISA's, then NVD's: High 7.6.
Weakness
Affected products
Capsule Technologies · Capsule Technologies SmartLinx Neuron 2
- affected 9.0.3 or lower
References (1)
- https://talosintelligence.com/vulnerability_reports/TALOS-2019-0785 (opens in a new tab)x_refsource_CONFIRMx_transferred
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket
[](https://trackcve.app/cve/CVE-2019-5024)Record last updated .