CVE-2019-6540
Medtronic Conexus Radio Frequency Telemetry Protocol Cleartext Transmission of Sensitive Information
- Published
- Updated
- Reserved
- Assigned by
- icscert
Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| icscertCNA (assigner) | 3.1 | Medium6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
| NVDNVD · Modified | 3.1 | Medium6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
- EPSS (exploitation in 30 days)
- 0.19% · top 91% of all CVEs (4 Nov 2026)
The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 6.5.
Weakness
Affected products
Medtronic · Conexus Radio Frequency Telemetry Protocol
- affected All versions
Medtronic · MyCareLink Monitor
- affected 24950
- affected 24952
- affected 2490C
Medtronic · CareLink 2090 Programmer
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
Medtronic · Protecta ICD, Protecta CRT-D
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
- affected All versions
What to do
Solution: Medtronic has developed mitigating patches for a subset of the affected implanted cardiac device models. These patches are installed during regular office visits. Medtronic has stated that patches for additional impacted models are being developed by Medtronic and will be deployed through future updates. Patches are currently available for the following affected models: * Protecta CRT-D and implantable cardioverter defibrillators (ICDs), all models * Amplia MRI CRT-D, all models (patch available in U.S. only) * Claria MRI CRT-D, all models (patch available in U.S. only) * Compia MRI CRT-D, all models (patch available in U.S. only) * Visia AF MRI ICD, all models * Visia AF ICD, all models * Brava CRT-D, all models * Evera MRI ICD, all models * Evera ICD, all models * Mirro MRI ICD, all models * Primo MRI ICD, all models * Viva CRT-D, all models
Workaround: Medtronic has applied additional controls for monitoring and responding to improper use of the Conexus telemetry protocol by the affected implanted cardiac devices. Additional mitigations are being developed and will be deployed through future updates, assuming regulatory approval. Medtronic recommends that users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: * Maintain good physical control over home monitors and programmers. * Use only home monitors, programmers, and implantable devices obtained directly from your healthcare provider or a Medtronic representative to ensure integrity of the system. * Do not connect unapproved devices to home monitors and programmers through USB ports or other physical connections. * Only use programmers to connect and interact with implanted devices in physically controlled hospital and clinical environments. * Only use home monitors in private environments such as a home, apartment, or otherwise physically controlled environment. * Report any concerning behavior regarding these products to your healthcare provider or a Medtronic representative. Medtronic has released additional patient focused information https://www.medtronic.com/security .
References (2)
- https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01 (opens in a new tab)x_refsource_MISCx_transferred
- 107544 (opens in a new tab)vdb-entryx_refsource_BIDx_transferred
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket
[](https://trackcve.app/cve/CVE-2019-6540)Record last updated .