CVE-2020-25152
B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
- Published
- Updated
- Reserved
- Assigned by
- icscert
Description
A session fixation vulnerability in the B. Braun Melsungen AG SpaceCom administrative interface Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows remote attackers to hijack web sessions and escalate privileges.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| icscertCNA (assigner) | 3.1 | Medium6.5 | CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N |
| NVDNVD · Modified | 3.1 | High8.1 | CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:N |
- EPSS (exploitation in 30 days)
- 1.3% · top 30% of all CVEs (4 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: no · Technical impact: partial
The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 6.5.
Weakness
Affected products
B. Braun Melsungen AG · SpaceCom
- affected unspecified – ≤ U61
- affected unspecified – ≤ L81
B. Braun Melsungen AG · Battery pack with Wi-Fi
- affected unspecified – ≤ U61
- affected unspecified – ≤ L81
B. Braun Melsungen AG · Data module compactplus
- affected A10
- affected A11
What to do
Solution: B. Braun recommends applying updates: SpaceCom: Version U62 or later (United States), L82 or later (outside the United States) Battery Pack SP with Wi-Fi: Version U62 or later (United States), L82 or later (outside the United States) Data module compactplus: Version A12 or later Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
Workaround: As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms: Ensure the devices are not accessible directly from the Internet. Use a firewall and isolate the medical devices from the business network. Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
References (2)
- https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02 (opens in a new tab)x_refsource_CONFIRMx_transferred
- https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html (opens in a new tab)x_refsource_CONFIRMx_transferred
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)