Skip to content
TrackCVE
  • Newest CVEsPage
  • Recently added to KEVPage
  • Rising exploitation (EPSS)Page
  • Weekly digestPage
  • StatisticsPage
  • WatchlistPage
  • API documentationPage
  • MCP serverPage
  • Data sourcesPage
  • ChangelogPage
  • AboutPage
  • How it's builtPage
Back

CVE-2020-25154

B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus

Medium5.4EPSS 0.66%
Published
Updated
Reserved
Assigned by
icscert

Description

An open redirect vulnerability in the administrative interface of the B. Braun Melsungen AG SpaceCom device Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to redirect users to malicious websites.

Timeline

  1. ID reserved

    By icscert

  2. · 20 months later

    Published

  3. · 3 years later

    Record updated

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2020-25154, by who scored it
Scored byCVSSScoreVector
icscertCNA (assigner)3.1Medium5.4CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N
NVDNVD · Modified3.1Medium6.1CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:C/C:L/I:L/A:N
EPSS (exploitation in 30 days)
0.66% · top 48% of all CVEs (7 Nov 2026)
CISA SSVC
Exploitation: none · Automatable: no · Technical impact: partial
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 5.4.

Weakness

Affected products

What to do

Solution: B. Braun recommends applying updates: SpaceCom: Version U62 or later (United States), L82 or later (outside the United States) Battery Pack SP with Wi-Fi: Version U62 or later (United States), L82 or later (outside the United States) Data module compactplus: Version A12 or later Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html

Workaround: As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms: Ensure the devices are not accessible directly from the Internet. Use a firewall and isolate the medical devices from the business network. Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html

References (2)

Sources

The same CVE in the primary databases and aggregators, to check against.

Badge for a README or a ticket

CVE-2020-25154 badge
[![CVE-2020-25154](https://trackcve.app/badge/CVE-2020-25154.svg)](https://trackcve.app/cve/CVE-2020-25154)

Record last updated .