CVE-2020-25166

B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus

High7.6EPSS 0.47%
Published
Updated
Reserved
Assigned by
icscert

Description

An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices.

Scores

CVSS scores of CVE-2020-25166, by who scored it
Scored byCVSSScoreVector
icscertCNA (assigner)3.1High7.6CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L
NVDNVD · Modified3.1High7.1CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:N/I:H/A:L
EPSS (exploitation in 30 days)
0.47% · top 59% of all CVEs (6 Nov 2026)
CISA SSVC
Exploitation: none · Automatable: no · Technical impact: partial
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: High 7.6.

Weakness

Affected products

What to do

Solution: B. Braun recommends applying updates: SpaceCom: Version U62 or later (United States), L82 or later (outside the United States) Battery Pack SP with Wi-Fi: Version U62 or later (United States), L82 or later (outside the United States) Data module compactplus: Version A12 or later Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html

Workaround: As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms: Ensure the devices are not accessible directly from the Internet. Use a firewall and isolate the medical devices from the business network. Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html