Skip to content
TrackCVE
  • Newest CVEsPage
  • Recently added to KEVPage
  • Rising exploitation (EPSS)Page
  • Weekly digestPage
  • StatisticsPage
  • WatchlistPage
  • API documentationPage
  • MCP serverPage
  • Data sourcesPage
  • ChangelogPage
  • AboutPage
  • How it's builtPage
Back

CVE-2021-41835

Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm

High7.3EPSS 0.30%
Published
Updated
Reserved
Assigned by
icscert

Description

Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on Port TCP/443, but the affected service does not perform an automated redirect from the unencrypted service on Port TCP/80 to the encrypted service.

Timeline

  1. ID reserved

    By icscert

  2. · 52 days later

    Published

  3. · 3 years later

    Record updated

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2021-41835, by who scored it
Scored byCVSSScoreVector
icscertCNA (assigner)3.1High7.3CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L
NVDNVD · Modified3.1High7.5CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS (exploitation in 30 days)
0.30% · top 77% of all CVEs (7 Nov 2026)
CISA SSVC
Exploitation: none · Automatable: yes · Technical impact: partial
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: High 7.3.

Weakness

Affected products

What to do

Solution: Fresenius Kabi has created new versions to address these vulnerabilities: • Link+ v3.0 (D16 or later) • VSS v1.0.3 (or later) • Agilia Connect Pumps Wifi Module (D29 or later) • Agilia Connect Partner v3.3.2 (or later) Fresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version. Fresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers’ installation, Fresenius Kabi recommends users to rely on CISA recommendation.

Workaround: CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: • Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet. • Locate control system networks and remote devices behind firewalls and isolate them from the business network. • When remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.

References (1)

Sources

The same CVE in the primary databases and aggregators, to check against.

Badge for a README or a ticket

CVE-2021-41835 badge
[![CVE-2021-41835](https://trackcve.app/badge/CVE-2021-41835.svg)](https://trackcve.app/cve/CVE-2021-41835)

Record last updated .