CVE-2022-43557
BD BodyGuard™ Pumps – RS-232 Interface Vulnerability
- Published
- Updated
- Reserved
- Assigned by
- BD
Description
The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.
Timeline
ID reserved
By BD
- 46 days later
· 46 days later
Published
- 2 years later
· 2 years later
Record updated
Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| BDCNA (assigner) | 3.1 | Medium5.3 | CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H |
| NVDNVD · Modified | 3.1 | Medium5.3 | CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H |
- EPSS (exploitation in 30 days)
- 0.23% · top 87% of all CVEs (7 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: no · Technical impact: partial
The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 5.3.
Weakness
Affected products
Becton, Dickinson and Company (BD) · BodyGuard™ Pump
- affected BD BodyGuard™
- affected CME BodyGuard™ 323 (2nd Edition)
- affected CME BodyGuard™ 323 Color Vision (2nd Edition)
- affected CME BodyGuard™ 323 Color Vision (3rd Edition)
- affected CME BodyGuard™ Twins (2nd Edition)
What to do
Workaround: BD recommends the following mitigations and compensating controls to reduce risk associated with this vulnerability: • Ensure physical access controls are in place and only authorized end-users have access to BD BodyGuard™ pumps. • Ensure that only BD-approved equipment is connected to the RS-232 interface of the affected pumps. • Ensure that no equipment is connected to the RS-232 interface when the affected pumps are delivering infusions. • Protect connected computer systems with BodyComm™ software with standard security measures.
References (1)
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket
[](https://trackcve.app/cve/CVE-2022-43557)Record last updated .