CVE-2023-0888
Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi
- Published
- Updated
- Reserved
- Assigned by
- B.Braun
Description
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| B.BraunCNA (assigner) | 3.1 | Medium4.9 | CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N |
| NVDNVD · Modified | 3.1 | High7.2 | CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 0.60% · top 51% of all CVEs (6 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: no · Technical impact: partial
The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 4.9.
Weakness
Affected products
B. Braun · Space Battery Pack SP with Wi-Fi
- affected ≤ 053L000092
- affected ≤ 054U000092
What to do
Solution: Update to version L93/U93 or newer solves the issue.
Workaround: Use secure credentials and keep them secret. Secure your network to prevent outside access to medical devices.
References (2)
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket