CVE-2024-48966
Life2000 service tools for test and calibration do not support user authentication
Critical10.0EPSS 0.69%
- Published
- Updated
- Reserved
- Assigned by
- Baxter
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| BaxterCNA (assigner) | 3.1 | Critical10.0 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
| NVDNVD · Deferred | 3.1 | Critical10.0 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 0.69% · top 47% of all CVEs (6 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: yes · Technical impact: total
Loading the EPSS history…
The list uses the assigner's score when there is one, then CISA's, then NVD's: Critical 10.0.
Weakness
Affected products
Baxter · Life2000 Ventilation System
- affected 06.08.00.00 and prior
References (1)
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket
[](https://trackcve.app/cve/CVE-2024-48966)Record last updated .