Back
CVE-2024-48973
Debug port on Life2000 Ventilator serial interface is enabled by default
Critical9.3EPSS 0.22%
- Published
- Updated
- Reserved
- Assigned by
- Baxter
Description
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| BaxterCNA (assigner) | 3.1 | Critical9.3 | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
| NVDNVD · Deferred | 3.1 | Critical9.3 | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 0.22% · top 87% of all CVEs (5 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: no · Technical impact: total
Loading the EPSS history…
The list uses the assigner's score when there is one, then CISA's, then NVD's: Critical 9.3.
Weakness
Affected products
Baxter · Life2000 Ventilation System
- affected 06.08.00.00 and prior