Back
CVE-2024-48974
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
Critical9.3EPSS 0.15%
- Published
- Updated
- Reserved
- Assigned by
- Baxter
Description
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| BaxterCNA (assigner) | 3.1 | Critical9.3 | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
| NVDNVD · Deferred | 3.1 | Critical9.3 | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 0.15% · top 95% of all CVEs (6 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: no · Technical impact: total
Loading the EPSS history…
The list uses the assigner's score when there is one, then CISA's, then NVD's: Critical 9.3.
Weakness
Affected products
Baxter · Life2000 Ventilation System